pH Cleaning
The following procedure applies specifically to
the IJ44 pH electrode. The solutions and procedures can also be adapted
for cleaning other styles of pH electrodes.
pH Cleaning
1. Clean the electrode periodically dependent on the application. Remove
the sleeve as in Preparation. Clean the membrane, ground
glass stem and sleeve with solvents, detergents, or acid. DO NOT USE ABRASIVE
MATERIALS!
-
Alkaline samples and scale - Soak the electrode in 6M HCl, 3M sulphuric,
or chromosulphuric acid for 15 minutes. Wash well with tap water and soak
before use.
-
Biological and protein samples - Soak the electrode in 0.1% pepsin / 0.1M
HCl for 1 hour or overnight. Rinse the electrode and then soak in pH7 buffer
which contains 0.5g KCl /100mL before use.
-
Grease and oil - Wipe the membrane and ground stem with cotton or tissue
soaked in acetone or methylated spirits. Wash with deionised water and
soak before use.
-
Dirt and organics - Wipe the tip with cotton or tissue soaked in a mild
non-alkaline detergent (e.g. Palmolive). Wash with deionised water and
soak before use.
2. Rejuvenation of the response of aged pH electrodes may be obtained by
immersing for 10 seconds (no longer) in 10% hydrofluoric acid and immediately
washing well in tap water. This treatment should be used sparingly.
pH Maintenance
When not in use, it is preferable to keep the electrode immersed in 2M
KCl. If this is not available, pH 4 or 7 buffer is acceptable. This keeps
the reference from drying out and the pH glass hydrated, so the electrode
is ready for use immediately.
How often does the electrolyte need replacing? This depends on the
sample being measured, and is best determined by experience. Error can
be determined by comparing readings before and after renewal of the electrolyte
on the same sample. Normally, errors of less than a few hundredths of a
pH can be expected after a full day's use and electrolyte need only require
weekly replacement.
For long term storage, remove and clean the sleeve. Replace the sleeve
without electrolyte and attach the wetting cap, filled with deionised water.
Note: After long term storage algae or mold
growth may occur inside the wetting cap. This can be removed by soaking
in dilute chlorine (bleach) solution.
pH Helpful Hints
For greatest accuracy in pH measurement, follow these guidelines:
-
Use the same technique to measure samples which was used for calibration.
Be consistent with stirring rates, times and conditions.
-
Calibrate with buffers which are close in temperature to that of the sample.
-
Calibrate the pH electrode regularly, e.g. once an hour for accuracy to
within 0.01 pH, or once a day for accuracy to within 0.1pH.
-
Use fresh buffers for calibration. Avoid contamination of the stock buffer
solution and do not use beyond the expiry date.
-
Use a dedicated pH meter. DO NOT USE A MULTIMETER!
-
Keep all connections dry.
-
Immerse the electrode far enough into the solution to insure the reference
junction (where the ground glass stem contacts the end of the sleeve) is
below the surface.
-
Allow adequate time for the electrode to stabilize in standards and samples
before taking a reading. Clean the electrode periodically. Allow more time
for aged electrodes.
-
Do not use the pH electrode in solutions of fluoride ion at low pH. This
will etch the glass membrane.
-
Sulphide vapors can permeate the electrode wick and contaminate the reference
element. Minimize contact in such environments and change the reference
electrolyte frequently.
-
This electrode is suitable for use in TRIS buffers, as the external reference
electrolyte does not contain silver ions which cause fouling.